120
SCAN-LAN VI
9. Press CONTINUE (F4) or ENTER. The message “CALIBRATING...”
will be displayed, and the bar graph across the bottom of the display
will show the progress of the calibration. Press CANCEL (F1) or ESC
to cancel the calibration procedure. When finished, the screen returns
to the CALIBRATE menu.
NOTE
If at any time during the above procedure an error message is displayed
on the screen, start at the beginning of the instructions and attempt the
calibration again. If the problem continues, contact Black Box for
technical support.
10.4 Impedance Calibration
The SCAN-LAN VI does not support impedance calibration.
10.5 Resistance Calibration
Resistance calibration should be performed on a periodic basis in order
to maintain maximum accuracy. Resistance calibration is most sensitive
to temperature changes. Allow the main unit to stabilize to the operating
environment before performing the calibration procedure. Calibration
should be performed whenever a different Performance Module is
installed.
To calibrate Resistance:
1. Connect the SCAN-LAN VI to the Remote Unit.
2. Set the dial to CAL.
3. Highlight Resistance in the Calibrate menu by using the arrow keys.
4. Press CALIBRATE (F4) or ENTER. The Calibrate Resistance screen
will appear. The bar graph across the bottom of the display will show
the progress of the calibration.
5. When finished, the screen returns to the CALIBRATE menu.
10.6 Main-Unit Calibration
This calibration mode is intended for factory use only.