Appendix B: Error Codes
General Errors,
Cont’d
Code
Cause
0201
NO_SENSOR_ERR
Carrier x-axis motor couldn’t find optical sensor –
occurs in a homing sequence during self-test, prior to
running a program, or following a program.
Probable Causes:
•
Carrier x-axis motor is unable to move due to
contamination of the carrier transport rail.
•
Carrier x-axis motor failed, belt is slipping, or
drive or idler pulley is defective.
•
Optical sensor is defective.
0202
NO_SENSOR_ERR
Magna models only:
Carrier x-axis motor couldn’t
find optical sensor – occurs in a homing sequence
during self-test, prior to running a program, or
following a program.
Probable Causes:
•
Carrier x-axis motor is unable to move due to
contamination of the carrier transport rail.
•
Carrier x-axis motor failed, belt is slipping, or
drive or idler pulley is defective.
•
Optical sensor is defective.
0203
NO_SENSOR_ERR
Dual-manifold models only
:
Carrier y-axis motor
couldn’t find optical sensor – occurs in a homing
sequence during self-test, prior to running a program,
or following a program.
Probable Causes:
•
Carrier y-axis motor is unable to move due to
contamination.
•
Optical sensor is defective.
•
Carrier y-axis motor is defective.
0300
AUTOCAL_JIG_ERR
Manifold motor couldn’t find edge of autocal jig during
autocal routine.
Probable Causes:
•
Distance from optical sensor to autocal jig is too
short.
•
Autocal jig contacts are contaminated.
•
Autocal jig is not installed properly or a
connection is not made.
Bio-Tek Instruments, Inc.
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...