Appendix B: Error Codes
Fatal Errors,
Cont’d
Code
Cause
A700
QFLASH_ERR
Quick flash memory read did not match write.
Occurs when verifying that configuration data
written to flash memory actually matches that
intended to be written, in size or content. This
could follow configuration downloads, saving of
autocal data or program files, etc.
A800
RAM_ERR
RAM error - occurs during failure of the RAM
test on the test station. (EPROM test station
error only.)
A900
HEAP_CORRUPTED_ERR
Memory manager (memory allocation heap)
corruption detected – occurs when writing data
to flash memory. This can also be detected
during that portion of the software, which runs
in the background, while other tasks are
simultaneously running.