Decontaminate the Washer
|
131
Washer not equipped with the
external valve module:
MAINT
SELECT MAINT PROGRAM:
DECONTAMINATION
CONNECT DISINFECTANT AND PRESS <
START
> KEY
RUNNING: DECONTAMINATION/ DISINFECTING
CONNECT RINSE BOTTLE AND PRESS <
START
> KEY
Washer equipped with the
external valve module:
MAINT
SELECT MAINT PROGRAM:
DECONTAMINATION
DISINFECTANT VALVE?
A B C
or
D
RINSE VALVE?
A B C
or
D
START MAINTENANCE?
YES