System Startup
|
45
System Startup
To turn on the ELx405
, press the on/off switch on the washer’s rear panel. The
washer will perform a System Self Test, displaying the screens shown below until
initialization is complete. During this time, all keys are inactive.
POWERUP SEQUENCE Vx.xx
INITIALIZING...
BIO-TEK ELx405
SELF-TEST.....
If the Self Test fails, the washer will “beep” and the display will show an error code.
Note the error code and then press the
STOP
key to stop the beeping. Refer to
Appendix A
,
Troubleshooting
and
Appendix B
,
Error Codes
, to determine its cause.
If the problem is something you can fix, turn off the washer, fix the problem, and then
turn the washer back on. If the cause is not something you can fix, contact Bio-Tek’s
Technical Assistance Center (see
Chapter 1
), for further assistance with
troubleshooting.