Linking Programs
|
89
Define Link Program
To define a Link program, follow the menu path shown below. Some of these items
are options and some are prompts. The options are described in the table below:
DEFINE SELECT ACTION CREATE SELECT PROGRAM TYPE
MORE
LINK
NAME
PLATE TYPE
SELECT LINK 01 TYPE
:
SELECT LINK 02 TYPE
:
SELECT LINK 03 TYPE
: (UP TO 10)
OK TO SAVE PROGRAM?
PLATE TYPE
only appears on the Select, Magna, Select CW, and HT models with
the accessory 96-tube manifold installed.
Define Link Program Options
Option & Description
Models
Action/Comments
Name
Enter a unique name for
the new Link program.
All
Use up to 16 alphanumeric characters.
See
Entering Program Names
on page
44 for instructions.
Select,
Select CW,
Magna,
HT/96-tube
Choose 96 or 384.
Note:
For higher throughput of the HT
with 384-well plates, use the 192-tube
manifold.
Plate Type
Specify the microplate
type.
Magna
Choose 96 or 96Bead.
Select Link (n) Type
Select each Wash, Prime,
Dispense, Aspirate,
and/or Soak program to
be included in the Link
program.
All
Programs will be run in the order in which
they are selected.
If the
Plate Type
option is presented
(see above), then only programs with the
selected Plate Type will be accessible for
inclusion in the Link program.
Press
Main Menu
to save
.
All
Select
YES
to save the program.
ELx405 Operator's Guide
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...