Appendix B: Error Codes
Fatal Errors
Fatal errors indicate conditions that require immediate attention. If a fatal error is
displayed, contact Bio-Tek’s Technical Assistance Center for further instructions.
For the codes that represent
device
errors, the fourth digit of the code represents the
affected device:
0
= manifold motor
1
= carrier motor (x-direction)
2
= magnet motor (Magna models)
3
= carrier motor (y-direction) (dual-manifold models)
4
= software timer
5
= display
6
= quick flash memory
7
= memory manager (memory allocation heap)
Code
Cause
A100
TCB_NOT_AVAIL_ERR
Task control block not available – occurs when
trying to create a new software task to run
“simultaneously” with other tasks, such as
when first trying to run a wash program, or
when trying to start an AUTOCLEAN program in
an Ultrasonic Advantage™ model.
A300
NOT_AVAIL_ERR
Manifold motor not available – occurs when
trying to get access to a system resource, such
as the manifold motor or a timer. If the
requested device is the manifold motor, usually
that motor is still running, or the software
“thinks” the motor is still running, even if it
isn’t.
A301
NOT_AVAIL_ERR
Carrier x-axis motor not available – occurs
when trying to get access to a system
resource, such as the carrier x-axis motor or a
timer. If the requested device is the x-axis
motor, usually that motor is still running, or
the software “thinks” the motor is still running,
even if it isn’t.
Bio-Tek Instruments, Inc.
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...