Appendix B: Error Codes
General Errors,
Cont’d
Code
Cause
0401
MOTOR_VERIFY_ERR
Carrier x-axis motor failed positional verify – occurs
when verifying the current motor position during self-
test, prior to running a program, or following a
program.
Probable Causes:
•
Carrier x-axis motor is unable to move due to
contamination of the carrier transport rail.
•
Carrier x-axis motor failed, belt is slipping, or
drive or idler pulley is defective.
•
Optical sensor is defective.
•
Incorrect manifold selected (HT2/HTV2 models).
0402
MOTOR_VERIFY_ERR
Magna models only:
Carrier x-axis motor failed
positional verify – occurs when verifying the current
motor position during self-test, prior to running a
program, or following a program.
Probable Causes:
•
Carrier x-axis motor is unable to move due to
contamination of the carrier transport rail.
•
Carrier x-axis motor failed, belt is slipping, or
drive or idler pulley is defective.
•
Optical sensor is defective.
0403
MOTOR_VERIFY_ERR
Dual-manifold models only
: Carrier y-axis motor
failed positional verify – occurs when verifying the
current motor position during self-test, prior to running
a program, or following a program.
Probable Causes:
•
Contamination on the y-axis rail of the carrier.
•
Y-axis motor is not able to move due to
contamination.
•
Optical sensor is defective.
•
Y-axis motor is defective.
•
Incorrect manifold selected (HT2/HTV2 models).
Bio-Tek Instruments, Inc.
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...