Wash
|
59
Define Wash DISPENSE (DISP) Program Options, Cont’d
Option & Description
Models
Action/Comments
Prime Volume
The volume of the buffer
or rinse used to prime the
dispense tubes.
All
The volume range is 1 to 999 ml.
Prime Flow Rate
The rate at which the
priming fluid is dispensed
from the tubes during the
prime.
All
The flow rate options range from 1 to 9,
where 1 = Slowest, 9 = Fastest.
Press
Main Menu
to save
All Select
YES
to save the program.
Wash DISPENSE (DISP) Defaults and Ranges
Parameter
Units
Default
Range
Models
50-3000
ELx405, Magna, Select,
Select CW, HT/96-tube
Dispense Volume
(In a Dispense-only
program, the default
volume is 250.)
µ
l/well 300
25-3000 HT/192-tube
1-9
All
Dispense Flow Rate
5
Additional
flow rates
10 and 11
Select CW
Vacuum on Volume
(Appears only in the
Select CW if Flow
Rates 10 or 11 were
selected.)
µ
l/well
10
50-3000
Select CW
Dispense Height
See note
below.
120 12-180
All
Notes:
•
Dispense Flow Rate - Select CW
: Flow rates 10 and 11
are not
recommended
for 384-well plates.
•
Dispense Height
: The display shows the corresponding measurement in mm.
Although the range is 12-180, the recommended highest value is
175
.
ELx405 Operator's Guide
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...