Liquid Tests
|
149
Test Solutions
SOLUTION #1: Buffer Solution
Pipette 1 ml Tween 20
into 1
liter (1000 ml) of deionized
water and mix well.
or
Pipette 10 ml of Bio-Tek Wetting Agent
(PN 7773002) into 1 liter of deionized
water and mix well.
SOLUTION #2: Evacuation Efficiency (Residual) Test Solution
Mix 100 ml of
Solution #1
with 0.050 grams of FD&C #1
blue dye.
or
Mix 90 ml of
Solution #1
with 10 ml of
Bio-Tek Blue Test Dye (PN 7773001)
.
SOLUTION #3: Dispense Precision Test Solution
(Single/Dual 96 manifolds)
Mix 1180 ml of deionized water with 20 ml of
Solution #2
.
SOLUTION #4: Dispense Precision Test Solution
(Dual 192 manifold)
Mix 1420 ml of deionized water with 20 ml of
Solution #2
.
You may use your own buffer solution in place of
Solution #1
. If
any tests fail using your own buffer, however, retry the tests using
the suggested solutions. Using pure DI water in place of Solution
#1 is
not
recommended and will likely result in the failure of the
unit to meet specifications.
Important:
The factory determined the pass/fail specifications for
the following tests using the test solutions described above.
Bio-Tek PN 7773002
(Solution #1 100X Concentrate Wetting
Agent 125 ml) contains 10% Tween 20 in deionized water and
0.01% Sodium Azide as a preservative.
Bio-Tek PN 7773001
(Solution #2 10X Concentrate Blue Test
Dye 125 ml) contains 5 g per liter FD&C Blue #1, 0.1% Tween 20
in deionized water, and 0.01% Sodium Azide as a preservative.
ELx405 Operator's Guide
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...