General Errors
|
189
General Errors,
Cont’d
Code
Cause
0301
AUTOCAL_JIG_ERR
Carrier x-axis motor couldn’t find edge of autocal jig.
Probable Causes:
•
Autocal jig contacts are contaminated.
•
Autocal jig is not properly installed or a
connection is not made.
0302
AUTOCAL_JIG_ERR
Magna models only:
Carrier x-axis motor couldn’t
find edge of autocal jig.
Probable Causes:
•
Autocal jig contacts are contaminated.
•
Autocal jig is not properly installed or a
connection is not made.
0303
AUTOCAL_JIG_ERR
Dual-manifold models only
: Carrier y-axis motor
couldn’t find edge of autocal jig.
Probable Causes:
•
Autocal jig contacts are contaminated.
•
Autocal jig is not properly installed or a
connection is not made.
0400
MOTOR_VERIFY_ERR
Manifold motor failed positional verify – occurs when
verifying the current motor position during self-test,
prior to running a program, or following a program.
Probable Causes:
•
Linear way is dirty, loosing steps due to lack of
grease, or an object is blocking its path.
•
Motor is defective.
•
Optical sensor is defective.
•
Shipping block is still installed.
•
Lead screw has become unglued from the motor
shaft.
•
Incorrect manifold selected (HT2/HTV2 models).
ELx405 Operator's Guide
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...