Appendix B: Error Codes
General Errors,
Cont’d
Code
Cause
0B00
MANIFOLD_AUTOCAL_ERR
Invalid manifold for calibration – occurs when the value
stored in flash memory for the manifold type is invalid.
This is checked at powerup, during self-test, and during
and after autocal is run.
Probable Causes:
•
Memory corruption.
0C00
INVALID_BUFFER_ERR
An invalid buffer is selected – occurs when a program
first starts, when fluid is going to be dispensed, or
when a linked program is going to the next link.
Probable Causes:
•
Memory corruption.
0E00
DISPENSE_RATE_ERR
Dispense rate is invalid for manifold type – occurs when
verifying validity of a program, e.g., during run-time, or
when saving a program.
Probable Causes:
•
Incorrect rate/volume entered for manifold
selected.
0F00
DISPENSE_VOLUME_ERR
Dispense volume invalid for manifold type – occurs
when verifying the validity of a program, e.g., during
run-time, or when saving a program.
Also occurs when attempting to dispense fluid, and the
transition volume for that rate is larger than the
volume to be dispensed. (Transition volume = the non-
linear volume dispensed when the dispense motor is
turned on and off.)
Probable Causes:
•
Incorrect rate/volume entered for manifold
selected.
1000
CNFG_DATA_ERR
Necessary configuration data is missing because flash
memory configuration data has not been properly
initialized, or flash memory is corrupted.
Probable Causes:
•
New basecode overwrites configuration values.
•
Unit has not been calibrated or lost calibration
values.
•
Main PCB has been replaced.
•
Memory corruption.
Bio-Tek Instruments, Inc.
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...