66
| Chapter 3: Operation
Wash ASPIRATE Defaults and Ranges, Cont’d
Parameter
Units
Default
Range
Models
Magnet Aspirate
Height
Magna only
0
-20 to 20
Magna (if 96bead
Plate Type was
selected)
-55 to 55
with 96-well
plates
Select, Select CW
HT/96-tube
Horizontal
Y Aspirate
Position
Select, Select CW,
HT only
quarter steps
0
-10 to 20
with 384-
well plates
Select, Select CW,
HT/96-tube,
HT/192-tube
Aspirate Rate
3
1 to 10
Aspirate Delay
milliseconds 0
0-5000
Crosswise
Aspirate?
No
No/Yes
All
Crosswise On
Final
Final, All
All
Crosswise
Height
24
12-180
All
-30 to 30
ELx405, Magna
-55 to 55
with 96-well
plates
Select, Select CW,
HT/96-tube
Crosswise
Horizontal
Position
quarter steps
0
-25 to 25
with 384-
well plates
Select, Select CW,
HT/96-tube
HT/192-tube
Notes:
•
The shaded options are
not
available for Aspirate-only programs.
•
Horizontal Y Aspirate Position
and
Crosswise Horizontal Position
: For
higher throughput of the HT with 384-well plates, use the 192-tube manifold.
•
Crosswise Height
: Although the range is 12-180, the recommended highest
value is
175
.
Bio-Tek Instruments, Inc.
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Page 250: ...220 Index Bio Tek Instruments Inc...