Troubleshooting Charts
|
183
Microplate Scratches
Problem
Possible Cause
Possible Solution
Microplate dispense or
aspiration height
adjustment too low.
Change the Dispense or Aspiration
Height in the program (see
Define
Wash DISPENSE or ASPIRATION
Program Options
in
Ch. 3
).
Scratches
on
microplate
bottom.
Microplate not properly
seated or strips not
level.
Reseat microplate in carrier or strips
in holder.
Make sure the carrier is clean (see
Cleaning the Plate Carrier System
in
Ch. 4
).
Try a different microplate or strip
holder. If the problem is unresolved,
the carrier may have to be realigned.
Contact Bio-Tek TAC.