148
| Chapter 5: Performance Verification/Qualification Tests
Materials
•
One clean, dry microplate per test to be performed:
Microplate Type
Liquid Tests
Evacuation Efficiency Test
Dispense Precision Test
Corning
®
Costar #3590
96-well plates
Dispense Precision Test, with Valve Module
Evacuation Efficiency Test (192)
Dispense Precision Test (192)
Corning Costar flat-
bottom 384-well plates
Dispense Precision Test (192), with Valve
Module
•
Precision balance with minimum capacity of 100 g and readability of 0.01 g
resolution
•
Pipettes and graduated beakers
•
Microplate absorbance reader capable of dual wavelength reading at
630/450 nm
•
Liquid Test Worksheets at the end of this chapter for recording data and
results
•
Deionized water
•
Test solutions:
Prepare these solutions, using
the recipes on the next page:
If this
manifold is
installed:
Sol. #1
Sol. #2
Sol. #3
Sol. #4
Single/96
or Dual/96
1 batch
(1000 ml)
1 batch
(100 ml)
1 batch
(1200 ml)
n/a
Dual/192
1 batch
(1000 ml)
1 batch
(100 ml)
n/a 1
batch
(1440 ml)
These volumes are sufficient for performing the standard and
supplemental Evacuation Efficiency tests, plus the Dispense
Precision test. In most cases, enough fluid will be left over to re-run
a test, if necessary.
If you will be performing the
annual OQ
with the
external valve
module
, you will need several additional liters of deionized water.
Bio-Tek Instruments, Inc.
Summary of Contents for ELx405
Page 2: ......
Page 4: ...ii Preface Bio Tek Instruments Inc...
Page 6: ...iv Preface Bio Tek Instruments Inc...
Page 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Page 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Page 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Page 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...