B00418AA
4-18
Processing Samples
4.6 Requisition for Routine and Emergency Samples
9
When selecting a profile, perform one of the following operations.
a. Select
Profile
to open the profile window and select a profile (or
multiple profiles).
b. Use the keyboard to enter a profile number into Profile and press
(
Enter
)
.
TIP
It is necessary to program profiles in
Menu List > Parameters >
Common Test Parameters > Profile > Sample
before a profile is
available to requisition. A maximum of 99 profiles can be created
for each sample type.
Each time a test is selected, the Selected Tests and Sample
Volume fields are updated.
10
Select the
Demographics
tab to open the Demographics window and
enter the patient information.
11
Confirm the entered information and select
Entry (F2)
. The entered
information is programmed.
12
Repeat steps 5 through 12 to requisition additional samples in the
same Sample Kind and Type. To change the Sample Kind or Type,
select
Exit (F1)
and repeat steps 2 through 12.
B00418AA_AU5800 QRG_EN.book Page 18 Monday, December 13, 2010 9:59 AM