B00418AA
6-16
As Needed Tasks
6.3 Program Calibrator Concentrations
4
Select
Set Conc Value (F5)
. The Set Conc Value window opens.
5
The concentration value(s) of the selected calibrator display. To
display or edit a different calibrator, select the calibrator name from
the "Calibrator" drop-down list.
6
Enter the concentration values (Conc) for each test (Test Name) in the
calibrator. The concentration values can only be entered for tests
programmed to the calibrator in
Calibration Specific
.
7
Repeat steps 5 and 6 for any other calibrator to verify or change the
concentration.
8
Select
Close
to close the Set Conc Value window.
9
If a calibrator concentration was changed, a confirmation message
appears in the Set Conc Value window. Select
OK
.
10
Select
Confirm (F1)
.