B00418AA
2-14
System Overview
2.4 Software Description
Message
Displays messages regarding system conditions that could affect analysis
results.
Color-coding indicates the level of the message:
If an appropriate message is selected, a window to instruct the operator in the
required action will be displayed. Selecting
OK
opens the appropriate menu to
perform the required corrective actions.
TIP
A red Message or highlight on the analyzer picture indicates a condition
that prevents starting that unit.
Red:
Analysis cannot be started on a unit until the message is
addressed.
Orange:
Analysis can be started. Review the message carefully and take
appropriate action.
Yellow:
Analysis can be started. Review the message carefully and take
appropriate action. The message can shift to Orange status (more
critical).
Green:
A notification of system status. There are no operational
problems.
B00418AA_AU5800 QRG_EN.book Page 14 Monday, December 13, 2010 9:59 AM