As Needed Tasks
6.9 Fix (Assign) a Reagent Position
1
From the
RB/CAL/QC
screen, select the samples and criteria to
transfer.
2
Select
Transfer to Online (F7)
to display the Transfer to Online
window.
3
Select
OK
. Online transfer is performed.
TIP
Select
Online Transfer Stop (F7)
to stop the transfer.
6.9 Fix (Assign) a Reagent Position
Fix (assign) reagents that are not barcoded. The analyzer does not read
barcodes on reagents in fixed positions.
Non-barcoded reagents must be placed in the assigned reagent position.
Reagent ID Read Errors only display in the
Details
tab when “Reagent
Display” is set to Position.
1
Select
Home > Reagent Management > Details
.
TIP
Reagent Status may have to be “checked” before a position can
be fixed. Select
Reagent Check (F5) > “Reset”
.
2
Select the Unit No.
3
Select
Position
from the "Reagent Display" drop-down list.
4
Select an available bottle position to fix the reagent.