As Needed Tasks
6.1 Program a New Test
7
Select
Calibration Specific
:
•
Select the test from the "Test Name" drop-down list. Select
Serum
,
Urine
,
Other-1
or
Other-2
from the "Type" drop-down
list.
•
Refer to the parameter sheet to determine if the Calibration Type
is “AB” or “MB”, and enter Calibration Specific parameters.
•
If the Calibration Type is “AB”:
—
Refer to the parameter sheet for Formula, Slope Check,
Factor Range, Allowable Range Check, Advanced
Calibration, Lot Calibration, and Stability.
—
Counts (replicates): Enter a number from 1 to 4. For more
information, refer to 4.5.3 Calibration Specific Menu in the
AU5800 User’s Guide.
—
Calibrator: Select the calibrator material.
—
Conc: Enter the calibrator concentration from the package
insert.
•
If the Calibration Type is “MB”:
—
Refer to the parameter sheet for Formula, Allowable Range
Check, Advanced Calibration, MB Type Factor, and
Stability.
—
Counts (replicates): Enter a number from 1 to 4. For more
information, refer to 4.5.3 Calibration Specific Menu in the
AU5800 User’s Guide.
—
The MB Type Factor must be entered for the Inner and Outer
cuvette wheel, and each unit(s). Select
Inner
or
Outer
from
the "Cuvette" drop-down list. Select
unit 1
,
2
,
3
, or
4
from
the "Unit No." option.
•
Verify the information and select
Confirm (F1)
.