B00418AA
5-7
Monitoring Results
5.5 Check for Error Flags and Alarms
5
5.5 Check for Error Flags and Alarms
5.5.1 Check for Error Flags
If a problem occurred during analysis, the system appends a flag to the analysis
results. It is important that the operator reviews each flag as it is generated and
identifies the root cause. No result should be reported with an unresolved,
unexpected flag.
If a test generated an error flag, the test name and error flag appear in red on
Sample Status
.
Check all generated results carefully for error flags and take the appropriate
action.
For a list of error flags and the flag definition, refer to
in chapter 8.
For details on error flags, refer to CHAPTER 9, Error Flags in the AU5800
User's Guide.
5.5.2 Check for Alarms
Check if any alarms occurred during analysis:
Select
Alarm List
(
on the lower, right corner of the screen).
B00418AA_AU5800 QRG_EN.book Page 7 Monday, December 13, 2010 9:59 AM