B00418AA
3-33
Daily Start Up
3.8 Start Analysis
3
2
Verify all units have a check in the box indicating the unit will be
started. If the box is gray and inaccessible, review the error list and
perform necessary corrective actions before starting. Remove the
check from the box if analysis is not required on a unit, and the unit
will not start.
3
Select
Edit Sample No.
to edit the starting sample number. This is
only necessary in Sequential analysis.
4
Select
Start
. The system initializes and analysis starts if no errors are
detected. The mode changes from Standby to Measure 1.
B00418AA_AU5800 QRG_EN.book Page 33 Monday, December 13, 2010 9:59 AM