B00418AA
3-28
Daily Start Up
3.7 Perform Quality Control (QC)
8
Select
Display Cup Set (F5)
to view the reagent blank, calibrator(s),
racks, and positions required for reagent blank and calibrator(s). Load
the reagent blanks and calibrators according to the list in the correct
blue and yellow racks. Select
Close
to close the window.
TIP
In
Display Cup Set (F5)
, the Volume (µL) is the sample volume
required based on the requisitioned tests. The dead volume is
not included.
If calibrator barcode operation is in use, the Rack No. in
Display
Cup Set (F5)
displays as 000x.
To requisition QC the same as the calibration requisition, select
QC Same Requisition (F4)
.
9
Load the racks on the rack set trays on the rack feeder unit. Load the
blue rack first, followed by the yellow rack(s) in numerical rack ID
order.
10
Select
Start
.
3.7 Perform Quality Control (QC)
Quality control analysis is used for confirming the system performance. It is
necessary for all diagnostic equipment.
Check the performance of the AU5800 regularly by analyzing QC samples.
Each laboratory should establish its own control frequency. QC samples should
be tested each time patient samples are tested and each time calibration is
performed. If any trends or sudden shifts in values are detected, review all
operating parameters.
Each laboratory should also establish guidelines to ensure corrective action if
controls do not recover within the specified limits.