B00418AA
3-26
Daily Start Up
3.6 Perform Calibration
2
Select
Serum
,
Urine
,
Other-1
or
Other-2
from the "Type" drop-down
list.
3
“Cuvette” defaults to display all tests assigned to the Inner and Outer
cuvette wheels. Select Inner or Outer to display only those tests
assigned to the Inner or Outer wheel.
4
“Unit No.” defaults to display tests assigned to All units. Select a unit
number to display only those tests assigned to that unit number.
5
If the analyzer automatic requisition is satisfactory, proceed to step 8.
TIP
Automatic reagent blank and calibration requisition occurs in the
following conditions:
•
Reagent blank or calibration Expired
•
Reagent blank or calibration Expired Soon
•
New bottle or lot number of reagent if Advanced Calibration is
in use
•
Reagent blank or calibration failed
Tests display with a blue box in the RB column when
requisitioned for reagent blank. Tests display with a blue box in
the RB column and a yellow box in the CAL column when
requisitioned for calibration.
Select
Auto Cal/QC Requisition (F3)
to reset the analyzer
automatic requisitions if changes were made to the requisition.
In
Reagent Management
, any of the
Reagent Check (F5)
options, including
Reset
, reset the calibration requisitions to the
analyzer automatic requisitions.