3-24
Daily Start Up
3.6 Perform Calibration
3.6 Perform Calibration
Requisition reagent blanks and calibrations. Automatic reagent blank and
calibration requisition occurs with the following conditions:
•
Reagent blank or calibration Expired
•
Reagent blank or calibration Expired Soon
•
New bottle or lot number of reagent if Advanced Calibration is in use
•
Reagent blank or calibration failed
Review the Comments in
Reagent Management > Status
to determine what
tests need to be calibrated.