B00418AA
5-8
Monitoring Results
5.5 Check for Error Flags and Alarms
“Level” shows the alarm level by numbers and colors.
The “Count” at the right top of the window indicates the number of alarms
within “Date Range”. 4,096 cases can be stored and displayed using the scroll
bar.
Level
Background Color
Extent
Level 1
Red
A fatal system abnormality exists.
Level 2
Yellow
An abnormality influencing the data exists.
Level 3
Green
There is no system abnormality.
The operation log is displayed.
B00418AA_AU5800 QRG_EN.book Page 8 Monday, December 13, 2010 9:59 AM