Daily Start Up
3.6 Perform Calibration
3
A one-point calibration correction can be requisitioned and
performed if the test is a multi-point calibration, and 1-Point
Calibration Point is programmed in
Calibration Parameters
.
If the system consists of only one unit, “Unit No.” is not
displayed.
6
Select
Start Entry (F1)
.
•
Select
Profile
to open the Profile window and select a profile,
then select
OK
.
•
To select a specific test(s), select from the RB or CAL column to
the right of the test name. Selecting from the RB column requisi-
tions only a reagent blank. Selecting from the CAL column requi-
sitions a reagent blank and calibration.
•
Select
Individual Requisition (F3)
to requisition sequenced bot-
tles (1, 2, 3, 4, 5) of the same test. Select a specific test and bottle
sequence by selecting the RB or CAL column,
Select All by Test
(requisitions all bottles of the selected test), or
Select All
(requi-
sitions all bottles for all tests). Select
Close
to register the requi-
sition, or
Cancel
to cancel the requisition.
TIP
Calibration profiles must be programmed in
Menu List >
Parameters > Common Test Parameters > Profile > RB/
Calibration
before a profile is available to requisition. A
maximum of 99 different profiles can be programmed, including
daily, weekly, and monthly calibration requirements.
A test must be programmed for Advanced Calibration in
Menu
List > Parameters > Calibration Parameters > Calibration
Specific
before it is available to requisition by bottle sequence
number in
Individual Requisition (F3)
.
7
Select
Entry (F2)
to register the requisition. Select
Exit (F1)
to cancel
the requisition.