B00418AA
3-9
Daily Start Up
3.4 ISE (Optional Unit) Start Up
3
Figure 3.1
For information about loading ISE reagents, refer to 8.19.5 Replacing
ISE Reagents in the AU5800 User's Guide, chapter 8.
TIP
There is a Buffer, MID Standard, and Reference reagent bottle
dedicated to each flowcell (Cell1 and Cell2).
If there are two flowcells and one flowcell runs out of reagent,
processing continues with the flowcell that has sufficient
reagent. Processing stops on the flowcell that ran out of reagent.
It is possible to replace a bottle of ISE reagent and perform an
ISE calibration when the analyzer is in a
Measure
mode. Use the
Auto Stop switch to reset Cell 1 or Cell 2 to
Ready
status, then
calibrate Cell 1 or Cell 2.
3
Close the front panel.
1.
MID Standard solution bottle (for Cell 1)
2.
Buffer solution bottle (for Cell 1)
3.
MID Standard solution bottle (for Cell 2)
4.
Buffer solution bottle (for Cell 2)
5.
Reference solution bottle (for Cell 1)
6.
Reference solution bottle (for Cell 2)
B00418AA_AU5800 QRG_EN.book Page 9 Monday, December 13, 2010 9:59 AM