As Needed Tasks
6.1 Program a New Test
6
•
Long Name: (Optional) Enter a maximum of 20 characters.
•
Reagent ID: Enter the first 3 digits of the reagent ID, or refer to
the parameter sheet for the reagent ID 3-digit code.
•
Alarm Shots: Enter the remaining test number to generate a
Reagent Short alarm. The default is 32.
•
Multi Reagent Switch: Select
Yes
. This allows the analyzer to
switch to a new set of R1/R2 when either the R1 or R2 of a set
becomes empty.
•
Cuvette: Select
Inner
,
Outer
, or
Both
.
•
Verify the information and select
Confirm (F1)
.
2
Select
Group of Tests
.
•
Select
Group 1
,
Group 2
, or
Group 3
from the "Group" drop-
down list.
•
Select
Edit (F1)
.
•
Select
Test Setting (F5)
.
•
Select the test to add to the Group. The test name highlights in
blue. Select
Close
.
•
To change the print order: Select a test to enable
Forward (F2)
and
Backward (F3)
. Move the test in the Group as needed to
change the print order.
•
Verify the information and select
Confirm (F1)
.
3
Select
Profile
.
•
Select
Edit (F1)
.
•
Select
Sample
,
RB/Calibration
, and
QC
as needed to add the test
to any required profile. Verify the Sample Type for each profile.
Note:
QC
default profiles are numbers 87-99, based on sample
type and Group.
•
Select a profile at the "Profile Name" drop-down list.
•
Select the test. The selected tests highlight in blue.
•
Verify the information and select
Confirm (F1)
.