B00418AA
3-31
Daily Start Up
3.7 Perform Quality Control (QC)
3
2
Select
Serum
,
Urine
,
Other-1
or
Other-2
from the "Type" drop-down
list.
3
“Cuvette” defaults to display all tests assigned to the Inner and Outer
cuvette wheels. Select Inner or Outer to display only those tests
assigned to the Inner or Outer wheel.
4
“Unit No.” defaults to display tests assigned to All units. Select a unit
number to display only those tests assigned to that unit number.
5
If the default QC profile displayed is satisfactory, proceed to step 8.
TIP
QC profiles must be programmed in
Menu List > Parameters >
Common Test Parameters > Profile > QC
before a profile is
available to requisition. A maximum of 99 different profiles can
be programmed, including the default profiles.
A test must be programmed for Advanced Calibration in
Menu
List > Parameters > Calibration Parameters > Calibration
Specific
before it is available to requisition by bottle sequence
number in
Individual Requisition (F3)
.
6
Select
Start Entry (F1)
.
a. Select
Profile
to open the Profile window and select a profile, then
select
OK
.
b. Select a test(s) to requisition QC (blue highlight) or de-select QC
(no blue highlight).
c. Select
Individual Requisition (F3)
to requisition sequenced
bottles (1, 2, 3, 4, 5) of the same test. Select a specific test and
bottle sequence,
Select All by Test
(requisitions all bottles of the
selected test), or
Select All
(requisitions all bottles for all tests).
Select
Close
to register the requisition, or
Cancel
to cancel the
requisition.
d. Select
Deselect All Tests (F6)
to deselect all tests. Select a test(s)
as required for QC.
B00418AA_AU5800 QRG_EN.book Page 31 Monday, December 13, 2010 9:59 AM