B00418AA
6-15
As Needed Tasks
6.3 Program Calibrator Concentrations
6
6.3 Program Calibrator Concentrations
Use this function to verify and change calibrator concentrations. Select the
calibrator name to verify and change the concentrations of all tests in a
calibrator from the same screen. This is the easiest way to change all
concentrations when the calibrator lot number changes.
CAUTION
Confirm calibrator concentration values in Calibration Specific.
IT IS CRITICAL ALL MULTI-POINT CALIBRATOR VALUES WERE
ENTERED CORRECTLY.
For details on calibration specific test parameters, refer to 4.5.3 Calibration
Specific Menu in the AU5800 User’s Guide, chapter 4.
1
Select
Menu List > Parameters > Calibration Parameters >
Calibrators
.
2
Select
Edit (F1)
. The window changes to the edit mode.
3
Select the calibrator “Name” to edit.
B00418AA_AU5800 QRG_EN.book Page 15 Monday, December 13, 2010 9:59 AM