3.
In the Create New Instrument Protocol dialog box (Figure 20), select an
application type: Sequencing, Fragment, or HID. The run module selection list is
filtered based on the application you select.
Figure 20
Create New Instrument Protocol
Note:
Normalization parameters circled in red are displayed for fragment
analysis and HID applications only.
4.
Specify settings (“Instrument protocol settings“ on page 163).
5.
Save the assay:
• If you are creating the assay from the Library, click Save.
• If you are creating the assay from the Assign Plate Contents screen, click
Apply to Plate or Save to Library.
Instrument protocol settings
Setting
Description
Application Type
• Sequencing
• Fragment analysis
• HID
Capillary Length, Polymer,
Dye set
Capillary length, polymer type, and dye set with which the protocol will be used
Run module
Factory-provided modules that specify instrument control parameters. For more
information, see Appendix B, “Run modules and dye sets“.
Protocol name
Name of the protocol. Names must be unique.
Chapter 7
Manage library resources
Instrument protocol library
7
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
163