spectral calibration using the Z dye set. Do not select Keep Spectral Calibration
Data.
3.
Select the number of wells and plate position in the instrument.
Note:
You do not create a plate in the software for the install check.
4.
Specify the starting well for the injection position in which you loaded the
standard in the plate.
Note:
If you navigate away from the Install Standard screen after you start the
install check, the starting well may be reset to A01. This is a display issue only;
the starting well you specify is used for the install check.
5.
(Optional) If you have not already run a spectral calibration, select Keep Spectral
Calibration Data to save the sequencing install standard run (if it passes) as a
spectral calibration.
Note:
The spectral calibration record will only be saved if the Keep Spectral
Calibration Data option is checked on the screen. If you uncheck the option,
create a separate spectral calibration from the Maintenance menu.
Chapter 6
Run calibrations and install checks
Run a Sequencing install check
6
124
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1