Prepare the instrument
1.
In the Dashboard, check consumables status (see “Check consumables status“ on
• Consumables are not expired
• Adequate injections remain for consumables
2.
Ensure that buffer levels are at the fill lines (see “Check buffer fill levels“ on
3.
Set the oven temperature, then click Start Pre-heat.
Pre-heat the oven and detection cell while you prepare for a run (detection cell
temperature is set by the software). Pre-heating helps mitigate subtle first-run
migration rate effects. The pre-heat function automatically turns off after 2 hours.
We recommend that you pre-heat the oven for at least 30 minutes before you
start a run if the instrument is cold.
Temperatures are displayed in red as they warm to the set-points. When
temperatures are at the set-point they are displayed in green. Temperatures may
fluctuate slightly when they reach the set-point as they stabilize.
4.
Check the pump assembly for bubbles and run the Remove Bubble wizard if
needed (see “Remove bubbles from the polymer pump“ on page 237).
Prepare the fragment/HID install check standard
Prepare the standard as described in the product insert. See Appendix D, “Catalog
• Fragment analysis: DS-33 GeneScan
™
Installation Standards with GeneScan
™
600
LIZ
™
Size Standard v2.0
• HID: AmpFℓSTR
™
Identifiler
™
Allelic Ladder (For HID install check)
Prepare the standard plate
IMPORTANT!
Do not use warped or damaged plates.
1.
Load the standards in any injection position in the plate. The example below
shows injection position 1, but you can specify the starting well for an injection
position.
IMPORTANT!
You do not create a plate in the software for the install check.
8-capillary
96-well plate
A1 through H1
Chapter 6
Run calibrations and install checks
Run a fragment/HID install check
6
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
131