47
PROFESSIONAL
INNOVATIVE BRANDING
SERVICE
Menu operation
This will bring up the list edit screen where the steps in the list can be added and edited.
Parameter ranges are:
Format,
including three options, Rectangle, S-Ramp and F-Ramp. Supports different quantity
parameters setting.
NO.,
List file NO.
Step Count,
set the total steps of the current list file, which setting range is from 1 through 8.
Save,
save edits to the memory when done.
Each step allows setting of the following parameters in F-Ramp format:
U/V,
voltage setting.
Usl/s,
slew rate for voltage.
I/A,
current setting.
Isl/s,
slew rate for current.
P/W,
power setting.
Ro/
Ω
,
programmable output impedance setting.
Wid/s,
width setting for this step.
Mode,
Cont(continuous) or Step (once).
Count,
repeat count.
When the List edit are complete, press the
Save
soft key to save all changes made to the list.
Press the [ESC] key turn to Load page.
Note:
Failure to Save a transient list will cause it to be lost when the power source is turned off.
46
Menu operation
PROFESSIONAL
INNOVATIVE BRANDING
SERVICE
5.9.1 PLS Testing
This menu allows the user to program the output sequence in the following mode:
●
List
Program
●
Step
●
5.9.1.1 List mode
The List mode is basic but the most versatile output mode. It allows a sequence of events to
be programmed as a list.
Press the
Edit
soft key to enter edit mode.
5.9 Output Mode
The Output Mode screen provides access to select special functions and built-in test modes.
Some of these functions are only available on the Professional version models.
The following Soft Keys are available in the Output Mode screen to access these functions:
●
PLS Testing
●
Special Func