1
6
Parallel/Series
Two or more this series power supplies can be configured in Master/Slave mode to increase
output current or voltage.
Parallel mode, up to 16 units, to increase output power & current. Series mode, up to 2 units,
to increase output voltage.
!
Warning!
DO NOT connect units in both series and parallel mode at the same time.
6.1 Configuring All Slave Units
Tap on Parallel/Series icon on Navigation page to enter Master/Slave setting menu. Select Slave
option and set the Slave No..
1 master and up to 15 slaves. For Series mode, there can only be one Slave.
Next, press the
Save
soft key to store the slave type selection setting to non-volatile memory.
The unit is now ready to be used in a Master/Slave configuration. Slave units will keep their
output OFF as they are controlled by the Master unit.
Parallel/Series
PROFESSIONAL
INNOVATIVE BRANDING
SERVICE
Menu operation
5.10 Language
Selection of the display language between English, Simplified Chinese or Traditional Chinese.
57
PROFESSIONAL
INNOVATIVE BRANDING
SERVICE
56
PROFESSIONAL
INNOVATIVE BRANDING
SERVICE
DO NOT connect different models in series or parallel mode. All units must be the same
model number.
DO NOT exceed the number of maximum units in series or parallel mode.
Set up all Slave units first, then set up the Master in Parallel/Series configurations.
There can only be ONE Master unit in any Master/Slave configuration.
Each Slave unit's number MUST be unique or the Master/Slave mode cannot be set.
In Current counting mode, set the basic output parameters according to test requirement.
Press [1]~[9] and [Enter] keys to set IL, which is the cut-off current for counting. Ib, the breaker
or fuse current.
After the setting, press the [On/Off] key to turn on the output. The count starts from when
output current has reached Ib setting and stop until the breaker or fuse is open, when the
output current is less than IL setting.