Calibration and Troubleshooting
7-4
M370430-01 Rev B
Step 1: Gain Calibration
Mea
s
u
re
d
V
a
lue
Programmed Value
Programmed Value
Figure 7-2
Calibration: Step 1 Gain Calibration
Adjust the gain so that the real line and ideal line intersect at a
programmed value of 90%.
Step 2: Offset Calibration
Mea
s
u
re
d
V
a
lue
Programmed Value
Programmed Value
Figure 7-3
Calibration: Step 2 Offset Calibration
Adjust the offset so that the real and ideal lines intersect at a
programmed value of 10%.
Mea
s
u
re
d
V
a
lue
M
eas
ure
d
V
al
ue