Calibration and Troubleshooting
M370430-01 Rev B
7-23
7
For example, you would type
:
CAL:
INP
:
ANAL:
CURR
0
.
4
3
if you read 0.43 volts from the voltmeter.
After performing offset calibration, it is highly recommended that
you repeat gain calibration.
SCPI Command
[:]CALibration[<channel>]:INPut:ANALog[:VOLTage]
{<voltmeter reading in volts>}
Calibrating the Input Current APG Signal
The input voltage APG calibration is necessary to ensure accurate
measurements when using the
:MEAS : APR?
and
:MEAS : APR: ISOL?
SCPI command queries to monitor the APG input signal.
Gain Calibration
To calibrate the CAP APG input gain calibration:
1.
Connect a short across the power supply output.
2.
Connect a voltmeter and a DC voltage source to the
non-isolated current APG input lines (J1. 10 and J1 .7).
3.
Turn the power supply On.
4.
Set the Current APG level to 4 V. Send the SCPI command:
SYST:COMM:APR:LEV:CURR 4
5.
Set the input source as Voltage analog APG. Send the SCPI
command:
SYST:REM:
SOUR:
CURR
AVOL
6.
Set the voltage source attached to the non-isolated current
APG input to 90% of the nominal current level, in this case
3.6 V.
7.
Turn the output power on.
8.
The power supply starts up in Constant Current mode.
9.
Read the input voltage value from the voltmeter connected in
step 2.
10.
Type the SCPI input current calibration command with
voltage noted in step 9 as the parameter “data”.
For example, you would type : CAL: INP: ANAL: CURR 3.49 if
you were reading 3.49 volts from the voltmeter.
Summary of Contents for Sorensen XG
Page 2: ......
Page 22: ...Contents xx M370430 01 Rev B ...
Page 96: ...Local Operation 3 48 M370430 01 Rev B Figure 3 5 Series Operation ...
Page 102: ...Local Operation 3 54 M370430 01 Rev B ...
Page 210: ...Remote Operation 5 76 M370430 01 Rev B ...
Page 276: ...Communications Options 6 66 M370430 01 Rev B ...
Page 378: ...Rack Mount Options and Installation Instructions D 12 M370430 01 Rev B ...
Page 381: ......