157
Repairing the Autosampler
Leak Sensor
Leak Sensor
Frequency
When defective
Tools required
None.
Parts required
Leak sensor 5061-3356.
Before beginning this procedure:
❏
Remove the top cover (page 128).
❏
Remove the transport assembly (page 132).
❏
Remove the sampling unit (page 134).
1
Disconnect the leak sensor from the SUD
board (labelled “leak sensor”). Remove the
sensor.
2
Feed the cable of the new sensor through the
base of the sampling unit to the SUD board.
3
Connect the connector to the connector
(labelled “leak sensor”)
.