28
Installing the Autosampler
Installing the Thermostatted Autosampler
Installing the Thermostatted Autosampler
1
Place the Thermostat on the bench.
2
Remove the front cover and route the condensation drain tube to the waste.
WA R N I N G
Make sure that the condensation tube is always above the liquid level in
the vessel. If the tube is located in liquid the condensed water cannot
flow out of the tube and the outlet is blocked. Any further condensation
will then remain in the instrument. This may damage the instruments
electronics.
Figure 7
Condensation Leak outlet
3
Install the LAN interface board in the sampler (if required), see “Agilent
1100 Series Interface Board to general purposes” on page 198.
4
Remove the adhesive tape which covers the front door.
5
Remove the front door and remove the transport protection foam.
Preparation
Locate bench space
Provide power connection
Unpack the Sampler and the Thermostat
Parts required
Sampler and Thermostat
Power cord, for the other cables see below and “Cable Overview” on page 185
Chemstation and/or Control Module G1323B.
Condensation drain tube
Waste bottle
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......