134
Repairing the Autosampler
Sampling Unit
Sampling Unit
Frequency
When accessing internal parts, or when defective.
Tools required
1/4 inch wrench (supplied in accessory kit). Flat-head screwdriver.
Parts required
Sampling unit G1313-60008
for G1313A
Sampling unit G1329-60008
for G1329A
Sampling unit G1329-60018
for G1389A
Sampling unit G2260-60008
for G2260A
The sampling units come without injection valve and analytical head assembly (see
“Sampling Unit Assembly” on page 162).
Before beginning this procedure:
❏
Remove the front cover.
❏
Remove the vial tray.
❏
Remove the top cover (page 128).
❏
Remove the transport assembly (page 132).
1
Remove the pump and column thermostat
capillaries (ports 1 and 6) from the injection
valve.
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......