97
Troubleshooting and Test Functions
Jittery (shaky) movement in X and or theta axes and/or when the needle goes
through the gripper arm into the via
Jittery (shaky) movement in X and or theta
axes and/or
when the needle goes through
the gripper arm into the via
With Error messages
❏
motor overtemp (0 or 2)
❏
movement failed (0 or 2)
N O T E
When a motor over temperature message has occurred, the sampler must be
turned off for about 10 minutes to allow the motor to cool down.
Step 1
Check the firmware and update to the latest revision if necessary.
Since firmware revision A.03.61 (resident A03.60) most of following errors
“movement failed”, “motor over temp” and “initialization failed (X-axis)” are
solved.
Step 2
Check the cleanliness of the transport rods (X-axis) and clean them.
If the rod is dirty or sticky, clean it with Isopropanol and wipe it with a lint
free cloth. The rod can be lubricated with the following synthetic oil: part
number 6040-0854.
N O T E
DO NOT use other lubricant as mentioned above.
Step 3
Lubricate the X-gear.
Friction can result in the belt slipping on the gear so that the position of the
belt teeth towards the gear changes.
To avoid this, apply some grease from the sample transport repair kit to the
X-motor-gear.
N O T E
Do not use other grease as the one in the kit and carefully follow the
instruction from the technical note.
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......