117
Repairing the Autosampler
Rotor Seal
Rotor Seal
C A U T I O N
The stator face is held in place by the stator head. When removing the stator
head, ensure the stator face does not fall out of the valve.
Frequency
Poor injection-volume reproducibility
Leaking injection valve
Tools required
1/4 inch wrench (supplied in accessory kit).
Hex key, 9/64 inch (supplied in accessory kit).
Parts required
Rotor seal 0100-1853 (Vespel)
for G1313A and G1329A
Rotor seal 0100-1849 (Tefzel)
for G1313A and G1329A
Rotor seal 0100-2088 (Vespel)
for G1389A
Rotor seal 0101-1268 (PEEK)
for G2260A
Before beginning this procedure:
❏
Remove the front cover.
1
Remove all capillary fittings from the
injection-valve ports.
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......