189
Parts and Materials
Remote Cables
Remote Cables
One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent 1100 Series
modules. The other end depends on the instrument to be connected to.
Agilent 1100 to 3390 Integrators
Connector
01046-60203
Pin
3390
Pin
Agilent 1100
Signal Name
Active
(TTL)
2
1 - White
Digital
ground
NC2 - Brown
Prepare run
Low
7
3 - Gray
Start
Low
NC4 - Blue
Shut down
Low
NC5 - Pink
Not
connected
NC6 - Yellow
Power on
High
NC7 - Red
Ready
High
NC8 - Green
Stop
Low
NC9 - Black
Start request Low
Summary of Contents for G1313A