128
Repairing the Autosampler
Top Cover and Foam
Top Cover and Foam
C A U T I O N
This procedure requires removal of the MIO-i
nterface
board. The board is
sensitive to electrostatic discharge. Always use the ESD strap when handling
electronic boards.
Frequency
When accessing internal parts.
Tools required
If interface board installed: Flat-head screwdriver.
Parts required
Foam kit G1313-68702
1
Switch off the autosampler at the main power
switch. Remove the power cable.
2
If installed, remove the ECB board.
3
Move the safety lever to the left.Lift the clips
on the top cover.
4
Lift the cover away
.
CellFrame
safety lever
clip
CellFrame