Valve to Mainpass Failed
Valve to Mainpass Failed
The injection valve failed to switch to the mainpass position.
The switching of the injection valve is monitored by two microswitches on
the valve assembly. The switches detect the successful completion of the
valve movement. If the valve fails to reach the mainpass position, or if the
microswitch does not close, the error message is generated.
Probable Causes
•
Defective injection valve.
•
Defective ASM board.
Suggested Actions
❏
Exchange the injection valve.
❏
Exchange the ASM board.