16
Installing the Autosampler
Site Requirements
Environment
Your autosampler will work within specifications at ambient temperatures
and relative humidity as described in Table 1.
C A U T I O N
Do not store, ship or use your autosampler under conditions where
temperature fluctuations may cause condensation within the autosampler.
Condensation will damage the system electronics. If your autosampler was
shipped in cold weather, leave it in its box, and allow it to warm up slowly
to room temperature to avoid condensation.
WA R N I N G
Using the autosampler at environmental temperatures higher than 50
°
C
(122
°
F) may cause the rear panel to become hot.
Table 1
Physical Specifications - Autosampler (G1313A / G1329A / G1389A / G2260A)
7\SH
6SHFLILFDWLRQ
&RPPHQWV
Weight
14.2 kg (31.3 lbs)
Dimensions
(height × width × depth)
200 × 345 × 435 mm
(8 × 13.5 × 17 inches)
Line voltage
100 – 120 or 220 – 240 VAC,
±
10 %
Wide-ranging capability
Line frequency
50 or 60 Hz,
±
5 %
Power consumption (apparent power)
Power consumption (active power)
300 VA
200 W
Maximum
Maximum
Ambient operating temperature
4 – 55
°
C (41 – 131
°
F)
Ambient non-operating temperature
-40 to 70
°
C (-4 to 158
°
F)
Humidity
< 95 %, at 25 – 40
°
C (77 – 104
°
F)
Non-condensing
Operating Altitude
Up to 2000 m (6500 ft)
Non-operating altitude
Up to 4600 m (14950 ft)
For storing the autosampler
Safety standards: IEC, CSA, UL
Installation Category II, Pollution Degree 2
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......