209
Introduction to the Autosampler
Sampling Unit
Injection-Valve
The two-position 6-port injection valve is driven by a stepper motor. Only
five of the six ports are used (port 3 is not used). A lever/slider mechanism
transfers the movement of the stepper motor to the injection valve. Two
microswitches monitor switching of the valve (bypass and mainpass end
positions).
No valve adjustments are required after replacing internal components.
Table 43
Injection-Valve Technical Data
Standard
Micro
Preparative MBB™
Motor type
4V, 1.2A stepper
motor
4V, 1.2A stepper
motor
4V, 1.2A stepper
motor
Seal material
Vespel™ (Tefzel™
available)
Vespel™
PEEK
Stator material
Ceramic/PEEK
Head coated SST
PEEK
Number of ports
6
6
6
Switching time
< 150 ms
< 150 ms
< 150 ms
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......