130
Repairing the Autosampler
Installing the Top Cover and Foam
Installing the Top Cover and Foam
Frequency
When accessing internal parts.
Tools required
If interface board installed: Flat-head screwdriver.
Parts required
None.
1
Install the top foam. Press the foam firmly
into place.
2
Engage the slot on the top plate onto the side
plate.
3
Fix the top plate in place with the three
screws.
4
Engage the top cover into the bottom cover.
CellFrame
CellFrame