155
Repairing the Autosampler
Power Supply
4
Lift the power supply out of the module.
5
Remove the coupler from the of the power
supply.
6
Insert the coupler onto the switch of the new
power supply.
7
Install the power supply in the module.
8
Connect the light pipe to the coupler.
9
Install the bottom foam. Ensure the cables
are positioned as shown.
CellFrame
CellFrame
CellFrame
CellFrame
CellFrame